Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Omega 869: (1 year) CAL VER /5520 DATE: 08-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 7 NUMBER OF LINES: 54 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # om869.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: 30 min. - REQUIRED 1.004 DISP Temperature :18 - 28 degC 1.004 DISP Humidity: <80% 1.005 HEAD INITIAL SETUP 1.006 DISP Remove the back cover. 1.007 PIC om869 1.008 DISP Connect the 5520A to the UUT as follows: 1.008 DISP [32] 5520A NORMAL HI to pin 1 and 4 of probe connector 1.008 DISP [32] 5520A NORMAL LO to pin 2 and 3 of probe connector 1.008 DISP 1.008 DISP Use the same type of wire for all connections. 1.009 HEAD {200degC RANGE} 1.010 DISP Select the 200degC range. 1.011 5520 -100.0degC 0.3U R1 4W 2.001 5520 0.0degC 0.3U R1 4W 3.001 5520 100.0degC 0.3U R1 4W 4.001 5520 195.0degC 0.3U R1 4W 5.001 HEAD {630degC RANGE} 5.002 DISP Select the 630degC range. 5.003 5520 220degC 1U R1 4W 6.001 5520 600degC 1U R1 4W 7.001 END